For the best results from your analysis with cNx Suite, you must acquire high-quality spectra. Pay particular attention to proper shimming, phase correction and baseline correction to produce spectra that you can accurately analyze and interpret using cNx Suite.
Using cNx Suite, you can quantitatively analyze spectra produced by NOESY pulse sequences (noesypr1d on Bruker instruments) with tmix of 100 ms. . You can use other pulse sequences for identification purposes, but absolute quantification based on alternate pulse sequences may be unreliable.
We do not recommand using gradient-based pulse sequences because they sometimes distort intense peaks at their base.
For the best cNx Suite analysis results, use the following parameters for NMR data collection:
-
Solvent = 90% H2O/10% D2O
-
pH = 7
-
Temperature = 25°C
-
Acquisition time = 4 s
-
Relaxation delay = 1 s
-
Spectral width = 12 ppm
-
Ghost (dummy) scans = 8
Tips and Tricks
-
cNx Suite can automatically detect DSS or TSP signals in most spectra. However, automatic detection relies on the water peak being centered in the spectrum. For best results, ensure that the water peak is centered in your spectra.