In this note we present a method for removing lineshape distortions from nuclear magnetic resonance (NMR) spectra prior to more detailed analysis. Reference deconvolution is a method of reconstructing an ideal spectrum by removing lineshape distortions caused by field inhomogeneity (see [14], [15]). Applying reference deconvolution has been found to improve the quality of fit of a spectrum using targeted profiling in Chenomx NMR Suite. However, successful application of reference deconvolution to a spectrum requires some preprocessing and the presence of an appropriate reference peak.

All experiments performed with NMR spectrometers are subject to a variety of distortions imposed by the instruments. The majority of these affect all signals in the same way. Reference deconvolution attempts to correct these distortions by using an experimental signal for a known resonance to compensate for the distortions imposed on the spectrum by the spectrometer.


The general method involves deconvoluting the whole experimental spectrum with the experimental lineshape of the reference signal, and reconvoluting it with the ideal lineshape of the reference signal. Practically, this is most efficiently applied to the time domain, by extracting the component of the experimental free induction decay (fid) arising from the reference signal, and multiplying the original fid by the complex ratio of the ideal and experimental fids for the reference signal [14].

Applying reference deconvolution to a spectrum assumes that all lineshapes in the spectrum are equally affected by field inhomogeneity. Thus, it is only effective in removing lineshape distortions that affect all signals in the spectrum equally. This includes the majority of distortions imposed on spectra by NMR spectrometers, including distortions due to imperfect static field homogeneity, changes in signal phase and amplitude due to pulse phase and flip angle errors, and frequency shifts due to changes in the static magnetic field [14].


The effectiveness of reference deconvolution in reconstructing a spectrum relies on the selection of an appropriate reference signal. Several factors contribute to the suitability of a signal for use as a reference. To avoid errors resulting from division by very small numbers, the decay envelope for the chosen signal should not drop to zero anywhere in the fid that is being processed. As a result, simple multiplets like symmetric doublets or triplets, or experimental lineshapes with significant doublet character due to "split fields" or poor shimming may not be good candidates for reference signals. The signal should also be well-separated from any other signals, to ensure that neither the absorption nor the dispersion components of the signal experience any overlap. Finally, the ideal form of the signal should be accurately defined throughout the relevant region of the spectrum, including any closely-associated satellite signals [14].

In general, many of the characteristics that make a good peak for use as a chemical shape indicator (CSI) also make a good reference signal for reference deconvolution. As a result, most spectra of samples containing common CSIs like 2,2-dimethylsilapentane-4-sulfonic acid (DSS), trimethylsilylpropionic acid (TSP) or tetramethylsilane (TMS) can benefit from this technique.

Reference deconvolution is a linear process, based on direct and inverse Fourier transforms. As a result, quantitative relationships in a spectrum are preserved when the transformation is applied. However, the best results for a given spectrum are only possible when the spectrum has been properly phased, baseline corrected and zero-filled prior to applying reference deconvolution [15]. Zero-filling should be done to at least one full power of two beyond the number of points acquired. For example, if 32K points (215) are acquired, zero-fill to 64K points (216), but if 33K points are acquired, zero-fill to 128K (217).

Reference deconvolution is an effective method of improving spectrum quality. It can serve the two major functions of correction and refinement when applied to experimental spectra. Correction refers to adjusting the experimental data to remove the influence of external, systematic distortions, such as shimming errors or static field inhomogeneities. Refinement refers to standardizing spectra across a dataset to establish a consistent basis for comparison of the individual spectra or analysis of the dataset as a whole.


To apply reference deconvolution to a spectrum using Chenomx NMR Suite, you must open the spectrum in the Processor module. For maximum effectiveness, the spectrum must be properly phased, baseline-corrected and zero-filled. Use CSI Edit mode to locate and approximate the shape of the CSI in the sample (typically DSS or TSP), then switch to reference deconvolution mode. Set the region size using the triangles along the horizontal axis, making sure to include the full CSI peak, but no other peaks (Figure 3). 13C satellites should be excluded, but if 29Si satellites are present, include them and check the Use DSS/TSP Satellites box. Finally, select a target linewidth and click the Accept button to apply the transformation.


Target linewidths in reference deconvolution can be used in two ways. Specifying a target linewidth value x Hz larger than the native linewidth will have the same effect as applying line broadening at the same value. Thus, there is no need to apply line broadening if reference deconvolution is to be used. A target linewidth smaller than the native linewidth will allow a degree of resolution enhancement, but such enhancement will be at the expense of the signal-to-noise ratio. If the target linewidth is equal to the native linewidth, only the lineshape correction effects mentioned previously will apply.

Targeted profiling involves comparison of an experimental spectrum to NMR spectral signatures of individual metabolites from a reference database [13]. Such comparison is most useful when the observed lineshapes in the experimental spectrum are as close to ideal as possible. Variations in lineshape introduced by different instruments, or different shimming techniques, may adversely affect the results of an analysis. Applying reference deconvolution to experimental spectra prior to analysis by targeted profiling can increase the consistency of the analysis from spectrum to spectrum by removing the effect of these systematic lineshape distortions. In some cases, spectra that would otherwise be unusable can be restored to a useful condition.


The lineshape of an internal reference can allow assessment of the shimming quality of a spectrum. In the absence of reference deconvolution, a poorly-shimmed urine spectrum (Figure 1) would be discarded as unusable, and would need to be reacquired. If the sample with which the spectrum was originally acquired were no longer available, the information contained in the spectrum would be lost. Applying reference deconvolution, however, corrects the shimming errors and allows proper analysis of the spectrum (see Figure 4 and Figure 5).


Creating a compound signature involves building a mathematical model of the compound based on an experimental spectrum. If an experimental spectrum is distorted by external effects, then models based on the spectrum will be inaccurate. Reference deconvolution can remove many of these distortions, bringing the experimental spectrum closer to its ideal shape (Figure 6).

Statistical analysis of NMR spectra using spectral binning (also known as "spectral bucketing") can also be affected by systematic lineshape distortions. Such distortions change the apparent frequency distribution throughout the spectrum, changing the integrated area in each bin. In some cases, this could result in artificial separations in principal component analysis (PCA), based on variables like shimming technique instead of true variation in the samples. Even if PCA results are not necessarily incorrect, subtle variations in frequency distributions can act as a confounding factor when interpreting such results. Reference deconvolution can help ensure that the observed variation among samples is inherent to the samples themselves.